Ultrasound Raytracing Simulation#

This example implements a standalone ultrasound raytracing simulator that generates realistic ultrasound images based on 3D meshes. The simulator uses Holoscan framework and DDS communication for realistic performance.

Quick Start#

Please ensure you have the NVIDIA OptiX Raytracing with Python Bindings to setup the raysim module correctly.

python -m simulation.examples.ultrasound_raytracing

Expected Behavior:

  • Terminal messages showing the progress of the simulation, e.g. [info] Timing Simulation took 1.389568 ms.

To visualize the ultrasound images, please check out the Visualization Utility, but it's not recommended to use it without moving the probe with robotic arm, which receives topic_ultrasound_info from the scripts such as sim_with_dds.py or Tele-op.

Configuration#

Optionally, the simulator supports customization through JSON configuration files. You need to create your configuration file following the structure below:

{
    "probe_type": "curvilinear",
    "probe_params": {
        "num_elements": 256,
        "sector_angle": 73.0,
        "radius": 45.0,
        "frequency": 2.5,
        "elevational_height": 7.0,
        "num_el_samples": 1,
        "f_num": 1.0,
        "speed_of_sound": 1.54,
        "pulse_duration": 2.0
    },
    "sim_params": {
        "conv_psf": true,
        "buffer_size": 4096,
        "t_far": 180.0
    }
}

Probe Parameters#

Parameter Description Default Value
num_elements Number of elements in the ultrasound probe 256
sector_angle Beam sector angle in degrees 73.0
radius Radius of the ultrasound probe in mm 45.0
frequency Ultrasound frequency in MHz 2.5
elevational_height Height of the elevation plane in mm 7.0
num_el_samples Number of samples in the elevation direction 1
f_num F-number (unitless) 1.0
speed_of_sound Speed of sound in mm/us 1.54
pulse_duration Pulse duration in cycles 2.0

Simulation Parameters#

Parameter Description Default Value
conv_psf Whether to use convolution point spread function true
buffer_size Size of the simulation buffer 4096
t_far Maximum time/distance for ray tracing (in units relevant to the simulation) 180.0

Minimal Configuration Example#

You only need to specify the parameters you want to change - any omitted parameters will use their default values:

{
    "probe_type": "curvilinear",
    "probe_params": {
        "frequency": 3.5,
        "radius": 55.0
    },
    "sim_params": {
        "t_far": 200.0
    }
}

Command Line Arguments#

Argument Description Default Value
--domain_id Domain ID for DDS communication 0
--height Input image height 224
--width Input image width 224
--topic_in Topic name to consume probe position topic_ultrasound_info
--topic_out Topic name to publish generated ultrasound data topic_ultrasound_data
--config Path to custom JSON configuration file with probe parameters and simulation parameters None
--period Period of the simulation (in seconds) 1/30.0 (30 Hz)
--probe_type Type of ultrasound probe to use ("curvilinear") "curvilinear"